Contact HPD

Model 103 Rainer ADR Cryostat


Model 103 Rainier adiabatic demagnetization refrigerator (ADR) cryostat is a compact cryostat that provides a surprisingly large experimental space: 26 cm diameter by 25 cm tall.

HPD’s attention to detail provides many high-quality cryogenic features, including cool down from 300 K to 4 K in just 15 hours, a vibration minimizing design, and 250 hours no-load regulation at 100 mK.

Furthermore, the Model 103 Rainier can also be reconfigured for your specific project.

Features

  • Cryomech PT407 Pulse Tube Cooler
  • Remote Rotary Valve Motor
  • Vibration isolation bellows
  • Electronically controlled motorized heat switch
  • Quick release flanges on vacuum jackets
  • Nickel plated aluminum thermal shields (50 K, 4 K)
  • HPD ADR (NIST design)
    • Kevlar Suspension
    • 1 K & 50 mK Stages
  • Optimized for versatility and flexibility
    • 4X NW50 ports on top flange
    • 2X NW25 ports on top flange
    • 1X NW63 port on vacuum jacket side
    • 1X Ø6″ port on vacuum jacket bottom
  • Robust internal construction
  • Custom Service Stand (Option)
    • Allows for transport & servicing
    • 76 cm of height adjustment

Specifications

  • Vacuum Jacket Size: Ø 35 cm X 69 cm tall
  • Experimental Volume: Ø 26 cm X 25 cm tall
  • Cool down from 300 K to 4 K in 15 hours
  • 1st Stage Cooling Power: 22 W @ 55 K
  • 2nd Stage Cooling Power: 0.6 W @ 4.2 K
  • GGG Cooling Capacity: 1.2 J @ 1 K
  • FAA Cooling Capacity: 120 mJ @ 100 mK
  • ADR Base Temperature: 50 mK
  • 250 hour no-load regulation at 100 mK
Download Specs
Request a Quote

Compare

Research Cryostat Models

Model 103
(ADR)

  1. Vacuum Jacket Size
    Ø 35 cm X 69 cm tall

  2. Experimental Volume
    Ø 26 cm X 25 cm tall

  3. Cool down from 300 K to 4 K in 15 hours

  4. 1st Stage Cooling Power
    22 W @ 55 K

  5. 2nd Stage Cooling Power
    0.6 W @ 4.2 K

  6. GGG Cooling Capacity
    1.2 J @ 1 K

  7. FAA Cooling Capacity
    120 mJ @ 100 mK

  8. ADR Base Temperature
    50 mK

  9. 250 hour no-load regulation at 100 mK

Request Quote

Model 102
(ADR)

  1. Vacuum Jacket Size
    33 cm X 22 cm X 66 cm tall

  2. Experimental Volume
    24 cm X 15 cm X 14 cm tall

  3. Cool down from 300 K to 4 K in 12 hours

  4. 1st Stage Cooling Power
    22 W @ 55 K

  5. 2nd Stage Cooling Power
    0.6 W @ 4.2 K

  6. GGG Cooling Capacity
    1.2 J @ 1 K

  7. FAA Cooling Capacity
    120 mJ @ 100 mK

  8. ADR Base Temperature
    50 mK

  9. 150 hour no-load regulation at 100 mk

View Model

Model 104
(ADR)

  1. Vacuum Jacket Size
    Ø 55 cm X 110 cm tall

  2. Experimental Volume
    Ø 44 cm X 60 cm tall

  3. Cool down from 300 K to 4 K in 16 hours

  4. 1st Stage Cooling Power 36 W @ 45 K*, 40 W @ 45 K+

  5. 2nd Stage Cooling Power 0.9 W @ 4.2 K*, 1.35 W @ 4.2 K+

  6. GGG Cooling Capacity
    1.2 J @ 1 K

  7. FAA Cooling Capacity
    120 mJ @ 100 mK

  8. ADR Base Temperature
    50 mK

  9. 150 hour no-load regulation at 100 mK

View Model

Model 106
(ADR)

  1. Vacuum Jacket Size
    Ø 44 cm X 65 cm tall

  2. Experimental Volume
    Ø 34 cm X 21 cm tall

  3. Cool down from 300 K to 4 K in 16 hours

  4. 1st Stage Cooling Power 22 W @ 55 K*, 36 W @ 45 K+, 40 W @ 45 K*

  5. 2nd Stage Cooling Power 0.6 W @ 4.2 K*, 0.9 W @ 4.2 K+, 1.35 W @ 4.2 K*

  6. GGG Cooling Capacity
    1.2 J @ 1 K

  7. FAA Cooling Capacity
    120 mJ @ 100 mK

  8. ADR Base Temperature
    50 mK

  9. 250 hour no-load regulation at 100 mK

View Model

Model 107
(ADR)

  1. Vacuum Jacket Size
    Ø 44 cm X 65 cm tall

  2. Experimental Volume
    Ø 34 cm X 20 cm tall

  3. 1st Stage Cooling Power 22 W @ 55 K*, 36 W @ 45 K+, 40 W @ 45 K*

  4. 2nd Stage Cooling Power 0.6 W @ 4.2 K*, 0.9 W @ 4.2 K+, 1.35 W @ 4.2 K*

  5. 3He Cooling Capacity
    6 J @ 1 K

  6. FAA Cooling Capacity
    200 mJ @ 100 mK

  7. ADR Base Temperature
    30 mK

  8. 300 hour no-load regulation at 100 mK

View Model

Model 105
(DR)

  1. Vacuum Jacket Size
    Ø 55 cm X 110 cm tall

  2. Experimental Volume
    Ø 38 cm X 19 cm tall

  3. 1st Stage Cooling Power
    40 W @ 45 K

  4. 2nd Stage Cooling Power
    1.5 W @ 4.2 K

  5. Lower Stages
    1.5 K, 0.6 K Still, 0.1 K Cold Plate

  6. Mixing Chamber Base Temperature
    30 mK

  7. Mixing Chamber Cooling Power
    200 µW @ 100 mK

View Model

*Optional pulse tube
+Default pulse tube

Compare

Research Cryostat Models

Model 103 (ADR)

  • Vacuum Jacket Size: Ø 35 cm X 69 cm tall
  • Experimental Volume: Ø 26 cm X 25 cm tall
  • Cool down from 300 K to 4 K in 15 hours
  • 1st Stage Cooling Power: 22 W @ 55 K
  • 2nd Stage Cooling Power: 0.6 W @ 4.2 K
  • GGG Cooling Capacity: 1.2 J @ 1 K
  • FAA Cooling Capacity: 120 mJ @ 100 mK
  • ADR Base Temperature: 50 mK
  • 250 hour no-load regulation at 100 mK

Model 102 (ADR)

  • Vacuum Jacket Size: 33 cm X 22 cm X 66 cm tall
  • Experimental Volume: 24 cm X 15 cm X 14 cm tall 
  • Cool down from 300 K to 4 K in 12 hours
  • 1st Stage Cooling Power: 22 W @ 55 K
  • 2nd Stage Cooling Power: 0.6 W @ 4.2 K
  • GGG Cooling Capacity: 1.2 J @ 1 K
  • FAA Cooling Capacity: 120 mJ @ 100 mK
  • ADR Base Temperature: 50 mK
  • 150 hour no-load regulation at 100 mk

Model 104 (ADR)

  • Vacuum Jacket Size: Ø 55 cm X 110 cm tall
  • Experimental Volume: Ø 44 cm X 60 cm tall
  • Cool down from 300 K to 4 K in 16 hours
  • 1st Stage Cooling Power: 36 W @ 45 K*, 40 W @ 45 K+
  • 2nd Stage Cooling Power: 0.9 W @ 4.2 K*, 1.35 W @ 4.2 K+
  • GGG Cooling Capacity: 1.2 J @ 1 K
  • FAA Cooling Capacity: 120 mJ @ 100 mK
  • ADR Base Temperature: 50 mK
  • 150 hour no-load regulation at 100 mK

*Optional pulse tube
+Default pulse tube

Model 106 (ADR)

  • Vacuum Jacket Size: Ø 44 cm X 65 cm tall
  • Experimental Volume: Ø 34 cm X 21 cm tall
  • Cool down from 300 K to 4 K in 16 hours
  • 1st Stage Cooling Power: 22 W @ 55 K*, 36 W @ 45 K+, 40 W @ 45 K*
  • 2nd Stage Cooling Power: 0.6 W @ 4.2 K*, 0.9 W @ 4.2 K+, 1.35 W @ 4.2 K*
  • GGG Cooling Capacity: 1.2 J @ 1 K
  • FAA Cooling Capacity: 120 mJ @ 100 mK
  • ADR Base Temperature: 50 mK
  • 250 hour no-load regulation at 100 mK

*Optional pulse tube
+Default pulse tube

Model 107 (ADR)

  • Vacuum Jacket Size: Ø 44 cm X 65 cm tall
  • Experimental Volume: Ø 34 cm X 20 cm tall
  • 1st Stage Cooling Power: 22 W @ 55 K*, 36 W @ 45 K+, 40 W @ 45 K*
  • 2nd Stage Cooling Power: 0.6 W @ 4.2 K*, 0.9 W @ 4.2 K+, 1.35 W @ 4.2 K*
  • 3He Cooling Capacity: 6 J @ 1 K
  • FAA Cooling Capacity: 200 mJ @ 100 mK
  • ADR Base Temperature: 30 mK
  • 300 hour no-load regulation at 100 mK

*Optional pulse tube
+Default pulse tube

Model 105 (DR)

  • Vacuum Jacket Size: Ø 55 cm X 110 cm tall
  • Experimental Volume: Ø 38 cm X 19 cm tall
  • 1st Stage Cooling Power: 40 W @ 45 K
  • 2nd Stage Cooling Power: 1.5 W @ 4.2 K
  • Lower Stages: 1.5 K, 0.6 K Still, 0.1 K Cold Plate
  • Mixing Chamber Base Temperature: 30 mK
  • Mixing Chamber Cooling Power: 200 µW @ 100 mK

Interested in our other ADR cryostats?

View More
Contact HPD